CoaguChek XS Instruction Manual: A Comprehensive Plan

This manual details the CoaguChek XS system’s procedures, encompassing safety, testing protocols, quality control, and data management.
It’s designed for operators involved in PT/INR monitoring,
ensuring accurate results and adherence to regulatory standards.

The system’s SAM program aims to improve testing quality and patient safety.
Refer to this guide for comprehensive instructions and troubleshooting assistance.
The CoaguChek XS is a portable, point-of-care (POC) device designed for the determination of prothrombin time (PT) and international normalized ratio (INR). This system facilitates convenient and rapid monitoring of oral anticoagulant therapy, primarily warfarin, enabling timely adjustments to medication dosages. It’s intended for use in various settings, including physician’s offices, anticoagulation clinics, and patient self-testing environments.
This system employs capillary blood sampling, typically obtained via a fingerstick, minimizing the need for venipuncture. The CoaguChek XS utilizes single-use test strips containing all the necessary reagents for the PT/INR analysis. The device features a user-friendly interface, displaying results in seconds, and incorporates quality control features to ensure accuracy and reliability.
Understanding the principles of coagulation testing and the significance of PT/INR values is crucial for effective utilization of the CoaguChek XS. Proper training and adherence to established protocols are essential to guarantee accurate results and optimal patient care. This manual provides detailed instructions on system operation, quality control procedures, troubleshooting, and maintenance, empowering users to confidently perform and interpret PT/INR tests.
The CoaguChek XS system is a valuable tool for managing patients on oral anticoagulation, contributing to improved patient outcomes and reduced risk of thromboembolic events.
Intended Use of the CoaguChek XS
The CoaguChek XS system is specifically intended for in vitro diagnostic use in the quantitative determination of prothrombin time (PT) and international normalized ratio (INR) in whole blood. It’s designed to aid in the monitoring of oral anticoagulant therapy, particularly warfarin sodium, and is a crucial component in managing patients at risk of thromboembolic events.

This system is intended for use by healthcare professionals in clinical settings, such as physician’s offices, hospital laboratories, and anticoagulation clinics. Furthermore, it is also cleared for patient self-testing, empowering individuals to actively participate in their own healthcare management under the guidance of their physician.
The CoaguChek XS is not intended for use in neonatal care or for making critical decisions requiring immediate medical intervention. It should not be used for diagnosing coagulation disorders, but rather for monitoring the effectiveness of anticoagulant treatment.

Proper training and adherence to the procedures outlined in this manual are essential for ensuring accurate and reliable results. The system is intended to be used with CoaguChek XS test strips only, and any attempt to use alternative reagents may compromise the accuracy of the test.
System Components and Overview
The CoaguChek XS system comprises several key components working in harmony to deliver accurate PT/INR results. These include the main instrument unit, featuring a user-friendly interface with a display screen and control buttons for easy operation. The system utilizes dedicated CoaguChek XS test strips, each containing the necessary reagents for the PT/INR assay.
A crucial element is the optical reading system, responsible for analyzing the blood sample on the test strip and generating the corresponding PT/INR value. The device requires a power adapter for operation and includes a port for data transfer via USB, enabling connectivity to a computer for data management and reporting.
Additionally, the system incorporates a built-in quality control feature, utilizing control solutions to verify the instrument’s functionality and the integrity of the test strips. The package also includes a lancing device for fingerstick blood collection, along with lancets and alcohol swabs for proper skin preparation.
Understanding the function of each component is vital for proper system operation, maintenance, and troubleshooting; Refer to the detailed diagrams and descriptions within this manual for a comprehensive overview of the CoaguChek XS system.
Safety Precautions in the Testing Area
Maintaining a safe testing environment is paramount when operating the CoaguChek XS system. Always adhere to universal precautions for handling blood and potentially infectious materials. Wear appropriate personal protective equipment (PPE), including gloves, at all times during the testing procedure to prevent exposure.
The testing area should be clean, well-lit, and free from clutter. Disinfect the work surface before and after each testing session using an approved disinfectant solution. Proper disposal of sharps, such as lancets and used test strips, is crucial; utilize designated sharps containers for safe containment.
Avoid touching the test strip’s reagent area to maintain its integrity and prevent contamination. If skin contact occurs, wash the affected area thoroughly with soap and water. Ensure adequate ventilation in the testing area.
Never eat, drink, or smoke in the testing area. Familiarize yourself with the Material Safety Data Sheets (MSDS) for all reagents and solutions used with the CoaguChek XS system. Report any accidental spills or exposures immediately to the appropriate personnel.
Preparing for Testing: Quality Control
Prior to performing any PT/INR tests with the CoaguChek XS system, rigorous quality control (QC) procedures are essential to ensure accurate and reliable results. Utilize commercially available control solutions – both low and normal – designed specifically for the CoaguChek XS.
Run QC checks at least once per day, or more frequently as dictated by your facility’s policies and procedures, and whenever a new lot of test strips is introduced. Record all QC results meticulously in a designated logbook or electronic system. Results must fall within the established acceptable range specified by the control manufacturer.
If QC results are outside the acceptable range, do not perform patient testing. Troubleshoot the issue by reviewing the testing procedure, checking the expiration dates of reagents and controls, and ensuring proper instrument function. Repeat the QC check after addressing potential issues.
Document any corrective actions taken and the subsequent QC results. Adherence to a consistent QC schedule and thorough documentation are vital for maintaining the integrity of the testing process and ensuring patient safety.
Fingerstick Blood Collection Procedure
Proper fingerstick technique is crucial for obtaining an adequate and representative blood sample for CoaguChek XS testing. Begin by selecting a finger, preferably the side of the fingertip, avoiding areas that are calloused or bruised. Warm the finger, if necessary, to promote blood flow.
Cleanse the puncture site thoroughly with an alcohol swab and allow it to air dry completely. Using a sterile lancet, perform a quick, firm puncture. Wipe away the first drop of blood with a clean gauze pad, as it may contain tissue fluid.
Gently massage the finger towards the puncture site to encourage the formation of a hanging drop of blood. Avoid excessive squeezing, which can dilute the sample with interstitial fluid. Apply the CoaguChek XS test strip to the blood drop, allowing it to draw in the required sample volume automatically.
Ensure adequate sample volume is collected to prevent errors. If insufficient blood is obtained, a new fingerstick may be necessary. Always adhere to universal precautions and dispose of lancets and contaminated materials properly.
Test Strip Handling and Storage

Proper handling and storage of CoaguChek XS test strips are paramount to ensure accurate and reliable PT/INR results. Test strips are individually sealed to protect them from moisture, temperature fluctuations, and light exposure – all factors that can compromise their integrity.
Always store test strips in their original container, tightly closed, in a cool, dry place, and away from direct sunlight. Avoid storing them in bathrooms or other humid environments. Do not refrigerate or freeze the test strips, as this can damage the reagents.
Before use, verify the expiration date printed on the test strip container. Do not use expired test strips, as they may yield inaccurate results. Remove a test strip from the container only immediately before performing a test. Avoid touching the test area of the strip with your fingers.
Once the container is opened, use the remaining test strips within the timeframe specified by the manufacturer (typically 3 months). Discard any unused test strips after this period. Proper storage and handling are essential for maintaining test strip performance.
Performing a PT/INR Test
To initiate a PT/INR test with the CoaguChek XS system, begin by ensuring the device is powered on and displaying the main menu. Insert a new test strip into the designated port, ensuring correct orientation as indicated by the arrows. The device will automatically recognize the strip and initiate the testing sequence.
Apply a sufficient drop of capillary blood to the sample port on the test strip, following the on-screen prompts. Avoid applying excessive blood, as this can lead to erroneous results. The system will draw the required amount of blood via capillary action.
Remain still during the 60-second countdown while the test is processing. Do not disturb the device or the test strip. Once the test is complete, the PT/INR result will be displayed on the screen, along with the patient’s ID (if entered) and the date/time of the test.
Record the result accurately in the patient’s record. If the result is outside the expected range, repeat the test using a new test strip and a fresh blood sample. Always follow established quality control procedures.
Understanding Test Results
The CoaguChek XS system displays PT/INR results numerically, representing the time it takes for blood to clot. PT (Prothrombin Time) measures the time in seconds for a clot to form, while INR (International Normalized Ratio) standardizes this time, allowing for comparison across different laboratories and reagents.
A normal INR range is typically between 0.8 and 1.2 for individuals not taking anticoagulants. Patients on warfarin or other vitamin K antagonists require a target INR range determined by their physician, often between 2.0 and 3.0, but this can vary.
Results below the target range suggest the blood is clotting too slowly, potentially increasing the risk of bleeding. Results above the target range indicate the blood is clotting too quickly, raising the risk of thrombosis (blood clots).

Always interpret results in conjunction with the patient’s clinical history and medication list. Communicate any significantly abnormal results to the healthcare provider immediately. The CoaguChek XS system may also display quality control flags; these must be addressed before relying on the test result.
Quality Control Procedures and Documentation
Regular quality control (QC) is crucial for ensuring the accuracy and reliability of CoaguChek XS test results. Perform QC checks at least daily, and whenever a new lot of test strips is used, or after instrument maintenance. Utilize commercially available control materials with known PT/INR values – both low and normal ranges are recommended.
Record all QC results meticulously in a designated logbook or electronic system. Document the date, time, control material lot number, and the obtained QC value. Compare the results to the manufacturer’s specified acceptable range. Any QC result falling outside this range indicates a potential problem with the system, test strips, or procedure.
If a QC failure occurs, immediately investigate the cause. Repeat the QC test; If the failure persists, do not use the system for patient testing. Troubleshoot according to the manual, and contact technical support if needed. Document all troubleshooting steps and corrective actions taken.
Maintain comprehensive records of all QC procedures, results, and any corrective actions. These records are essential for demonstrating compliance with regulatory requirements and accreditation standards.
Troubleshooting Common Errors
“Err 101 – Test Strip Error” often indicates improper strip insertion or a damaged strip. Ensure the strip is inserted correctly, with the gold contacts facing the instrument. Try a new strip from a different lot. “Err 102 – Sample Error” suggests insufficient sample volume or air bubbles. Repeat the fingerstick, ensuring a proper drop is formed and applied to the test strip.
If you encounter “Err 103 – Instrument Error,” restart the device. If the error persists, contact Roche technical support. “PT/INR value out of range” could stem from improper technique, patient factors, or a faulty strip. Review the procedure, consider patient variables, and use a new strip.

For connectivity issues, verify the USB cable connection and software installation. Ensure the CoaguChek XS is compatible with your computer’s operating system. If the device won’t power on, check the power adapter and outlet.
Always consult the full error code list in the manual for detailed explanations and solutions. Document all errors and troubleshooting steps taken for future reference and potential support requests.
Maintenance and Cleaning
Regular maintenance ensures accurate results and prolongs the CoaguChek XS system’s lifespan. After each test, discard the used test strip and lancet safely in a sharps container. Wipe the sample port with a clean, lint-free cloth dampened with 70% isopropyl alcohol. Avoid harsh chemicals or abrasive cleaners, as they can damage the device.
Weekly, inspect the device for any physical damage, such as cracks or loose parts. Clean the exterior surfaces with a damp cloth. Monthly, perform a quality control check using control materials to verify the system’s accuracy. Document all quality control results.
Do not immerse the device in liquid. Ensure the device is completely dry before use. Periodically check the power adapter and cable for damage. Replace if necessary. Follow Roche’s recommended maintenance schedule outlined in the full instruction manual.
Proper cleaning and maintenance are crucial for maintaining optimal performance and preventing contamination. Always refer to the manual for specific cleaning instructions and safety precautions.
Data Management and Connectivity
The CoaguChek XS system offers robust data management capabilities, allowing for efficient tracking of PT/INR results. Test results are stored internally and can be downloaded to a computer using the CoaguChek XS software and a USB connection. This software enables data analysis, trend monitoring, and report generation.
Data can be exported in various formats, facilitating integration with Electronic Health Records (EHR) and other healthcare information systems. Ensure data security and patient privacy by implementing appropriate access controls and data encryption measures. Regular data backups are recommended to prevent data loss;
The system supports connectivity to laboratory information systems (LIS), streamlining workflow and reducing manual data entry. Connectivity options may vary depending on the specific system configuration. Refer to the system’s technical documentation for detailed instructions on data transfer and integration.
Proper data management is essential for maintaining accurate patient records and supporting informed clinical decision-making. Always adhere to relevant data privacy regulations and institutional policies.

SAM Program Purpose and Implementation
The SAM (Systematic Assessment of Monitoring) Program for the CoaguChek XS is a crucial component designed to ensure the accuracy and reliability of PT/INR testing performed outside of a traditional laboratory setting. Its primary purpose is to enhance the quality of anticoagulation management and improve patient safety by verifying operator competency and system performance.
Implementation involves regular assessment of testing procedures, including quality control checks, adherence to standardized protocols, and documentation of results. Operators participate in proficiency testing to demonstrate their ability to obtain accurate and consistent results. The program also includes ongoing training and education to maintain operator skills and knowledge.
Successful completion of the SAM Program is often a requirement for continued operation of the CoaguChek XS system in point-of-care settings. Detailed guidelines and resources are available through Roche Diagnostics to support program implementation and compliance. Regular audits and reviews are conducted to ensure program effectiveness.
Adherence to the SAM Program is vital for maintaining the integrity of PT/INR monitoring and optimizing patient outcomes.
Regulatory Compliance and Accreditation

The CoaguChek XS system, when utilized for patient care, is subject to various regulatory requirements and accreditation standards. These regulations are designed to ensure the accuracy, reliability, and safety of PT/INR testing. Compliance with these standards is essential for healthcare facilities employing the system.
Key regulatory bodies, such as CLIA (Clinical Laboratory Improvement Amendments) in the United States, establish guidelines for laboratory operations, quality control, and personnel qualifications. Accreditation organizations, like CAP (College of American Pathologists) and Joint Commission, offer voluntary accreditation programs that demonstrate a facility’s commitment to quality and adherence to best practices.
The CoaguChek XS system must be operated in accordance with manufacturer’s instructions and relevant regulatory guidelines. Proper documentation of quality control procedures, operator training, and maintenance activities is crucial for demonstrating compliance during inspections and audits. Regular participation in proficiency testing programs is also often required.
Maintaining regulatory compliance and accreditation status is vital for ensuring patient safety, reimbursement eligibility, and the continued operation of the CoaguChek XS system within a healthcare setting.